Data quality assessment is an integral part of preparatory data analysis to ensure sound biological information retrieval.
We present here the MsQuality
package, which provides functionality to
calculate quality metrics for mass spectrometry-derived, spectral data at the
per-sample level. MsQuality
relies on the
mzQC
framework of quality metrics defined
by the Human Proteome Organization-Proteomics Standards Intitiative (HUPO-PSI).
These metrics quantify the quality of spectral raw files using a controlled
vocabulary. The package is especially addressed towards users that acquire
mass spectrometry data on a large scale (e.g. data sets from clinical settings
consisting of several thousands of samples): while it is easier to control
for high-quality data acquisition in small-scale experiments, typically run
in one or few batches, clinical data sets are often acquired over longer
time frames and are prone to higher technical variation that is often
unnoticed. MsQuality
tries to address this problem by calculating metrics that
can be stored along the spectral data sets (raw files or feature-extracted
data sets). MsQuality
, thus, facilitates the tracking of shifts in data
quality and quantifies the quality using multiple metrics. It should be thus
easier to identify samples that are of low quality (high-number of missing
values, termination of chromatographic runs, low instrument sensitivity, etc.).
We would like to note here that these metrics only give an indication of
data quality, and, before removing indicated low-quality samples from the
analysis more advanced analytics, e.g. using the implemented functionality
and visualizations in the MatrixQCvis
package, should be scrutinized.
Also, data quality
should always be regarded in the context of the sample type and experimental
settings, i.e. quality metrics should always be compared with regard to the
sample type, experimental setup, instrumentation, etc..
The MsQuality
package allows to calculate low-level quality metrics that
require minimum information on mass spectrometry data: retention time,
m/z values, and associated intensities.
The list included in the mzQC
framework is excessive, also including
metrics that rely on more high-level information, that might not be readily
accessible from .raw or .mzML files, e.g. pump pressure mean, or rely
on alignment results, e.g. retention time mean shift, signal-to-noise ratio,
precursor errors (ppm).
The MsQuality
package is built upon the Spectra
and the MsExperiment
package.
Metrics will be calculated based on the information stored in a
Spectra
object and the respective dataOrigin
entries are used to
distinguish between the mass spectral data of multiple samples.
The MsExperiment
serves as a container to
store the mass spectral data of multiple samples. MsQuality
enables the user
to calculate quality metrics both on Spectra
and MsExperiment
objects.
MsQuality
can be used for any type of experiment that can be represented as
a Spectra
or MsExperiment
object. This includes simple LC-MS data, DIA or
DDA-based data, ion mobility data or MS data in general. The tool can thus
be used for any type of targeted or untargeted metabolomics or proteomics
workflow. Also, we are not limited to data files in mzML format, but, through
Spectra
and related MsBackend
packages, data can be imported from a large
variety of formats, including some raw vendor formats.
In this vignette, we will (i) create some exemplary Spectra
and MsExperiment
objects, (ii) calculate the quality metrics on these
data sets, and (iii) visualize some of the metrics.
Other R
packages are available in Bioconductor that are able to assess
the quality of mass spectrometry data:
artMS
uses MaxQuant output and enables to calculate several QC metrics, e.g.
correlation matrix for technical replicates, calculation of total sum of
intensities in biological replicates, total peptide counts in biological
replicates, charge state distribution of PSMs identified in each biological
replicates, or MS1 scan counts in each biological replicate.
MSstatsQC
and the visualization tool
MSstatsQCgui
require csv files in long format from spectral processing tools such as
Skyline and Panorama autoQC or MSnbase
objects. MSstatsQC
enables to
generate individual, moving range, cumulative sum for mean, and/or
cumulative sum for variability control charts for each metric. Metrics
can be any kind of user-defined metric stored in the data columns for a given
peptide, e.g. retention time and peak area.
MQmetrics
provides a pipeline to analyze the quality of proteomics data sets from
MaxQuant files and focuses on proteomics-/MaxQuant-specific metrics, e.g.
proteins identified, peptides identified, or proteins versus
peptide/protein ratio.
MatrixQCvis
provides an interactive shiny-based interface to assess data quality at various
processing steps (normalization, transformation, batch correction, and
imputation) of rectangular matrices. The package includes several diagnostic
plots and metrics such as barplots of intensity distributions, plots to
visualize drifts, MA plots and Hoeffding's D value calculation, and
dimension reduction plots and provides specific tools to analyze data sets
containing missing values as commonly observed in mass spectrometry.
proBatch
enables to assess batch effects in (prote)omics data sets and corrects these
batch effects in subsequent steps. Several tools to visualize data quality are
included in the proBatch
packages, such as barplots of intensity
distributions, cluster and heatmap analysis tools, and PCA dimension
reduction plots. Additionally, proBatch
enables to assess diagnostics at
the feature level, e.g. peptides or spike-ins.
To install this package, start R
and enter:
if (!requireNamespace("BiocManager", quietly = TRUE)) install.packages("BiocManager") if (!requireNamespace("remotes", quietly = TRUE)) install.packages("remotes") ## to install from Bioconductor BiocManager::install("MsQuality") ## to install from GitHub BiocManager::install("tnaake/MsQuality")
This will install this package and all eventually missing dependencies.
MsQuality
is currently under active development. If you discover any bugs,
typos or develop ideas of improving MsQuality
feel free to raise an issue
via GitHub or send a mail to the
developer.
Spectra
and MsExperiment
objectsLoad the Spectra
package.
library("Spectra") library("MsExperiment") library("MsQuality")
Spectra
and MsExperiment
objects from mzML filesThere are several options available to create a Spectra
object. One way, as
outlined in the vignette of the r BiocStyle::Biocpkg("Spectra")
package is
by specifying the location of mass spectrometry raw files in mzML
, mzXML
or
CDF
format and using the MsBackendMzR
backend. Here we load the example
files from the sciex
data set of the msdata
package and create a Spectra
object from the two provided mzML
files. The example is taken from the
Spectra
vignette.
## this example is taken from the Spectra vignette fls <- dir(system.file("sciex", package = "msdata"), full.names = TRUE) sps_sciex <- Spectra(fls, backend = MsBackendMzR())
The data set consists of a single sample measured in two different
injections to the same LC-MS setup. An empty instance of an
MsExperiment
object is created and populated with information on the samples
by assigning data on the samples (sampleData
), information on the
mzML
files (MsExperimentFiles
) and spectral information (spectra
).
In a last step, using linkSampleData
, the relationships between the samples
and the spectral information are defined.
## this example is taken from the Spectra vignette lmse <- MsExperiment() sd <- DataFrame(sample_id = c("QC1", "QC2"), sample_name = c("QC Pool", "QC Pool"), injection_idx = c(1, 3)) sampleData(lmse) <- sd ## add mzML files to the experiment experimentFiles(lmse) <- MsExperimentFiles(mzML_files = fls) ## add the Spectra object to the experiment spectra(lmse) <- sps_sciex ## use linkSampleData to establish and define relationships between sample ## annotations and MS data lmse <- linkSampleData(lmse, with = "experimentFiles.mzML_file", sampleIndex = c(1, 2), withIndex = c(1, 2))
Spectra
and MsExperiment
objects from (feature-extracted) intensity tablesAnother common approach is the creation of Spectra
objects from a
DataFrame
s using the MsBackendDataFrame
backend.
We will use here the data set of @Lee2019, that contains metabolite level
information measured by reverse phase liquid chromatography (RPLC) coupled
to mass spectrometry and hydrophilic interaction liquid chromatography (HILIC)
coupled to mass spectrometry
(derived from the file STables - rev1.xlsx
in the Supplementary Information).
In a separate step (see documentation for Lee2019_meta_vals
and
Lee2019
), we have created a list containing Spectra
objects for
each samples (objects sps_l_rplc
and sps_l_hilic
) and MsExperiment
objects containing the data of all samples (objects msexp_rplc
and
msexp_hilic
). We will load here these objects:
data("Lee2019", package = "MsQuality")
The final data set contains r length(unique(sps_rplc$dataOrigin))
paired samples (i.e. r length(unique(sps_rplc$dataOrigin))
samples derived
from RPLC and r length(unique(sps_hilic$dataOrigin))
samples derived
from HILIC).
We will combine the sps_rplc
and sps_hilic
objects in the following and
calculate on this combined document the metrics.
sps_comb <- c(sps_rplc, sps_hilic)
The most important function to assess the data quality and to calculate the
metrics is the calculateMetrics
function. The function takes
a Spectra
or MsExperiment
object as input, a character vector of metrics
to be calculated, and, optionally a list of parameters passed to the
quality metrics functions.
Spectra
and MsExperiment
objectsCurrently, the following metrics are included:
qualityMetrics(sps_comb)
The following list gives a brief explanation on the included metrics. Further
information may be found at the
HUPO-PSI mzQC project page or in the
respective help file for the quality metric (accessible by e.g. entering
?chromatographyDuration
to the R console).
We also give here explanation on how the metric is calculated in MsQuality
.
Currently, all quality metrics can be calculated for both Spectra
and
MsExperiment
objects.
The metric is calculated as follows:
(1) the retention time associated to the Spectra
object is obtained,
(2) the maximum and the minimum of the retention time is obtained,
(3) the difference between the maximum and the minimum is calculated and returned.
- rtOverMsQuarters, MS1 quarter RT fraction (MS:4000055),
"The interval used for acquisition of the first, second, third, and fourth
quarter of all MS1 events divided by retention time duration." [PSI:MS],
msLevel = 1L
;
The metric informs about the dynamic range of the acquisition along the
chromatographic separation. For MS1 scans, the values are expected to be in
a similar range across samples of the same type.
The metric is calculated as follows:
(1) the retention time duration of the whole Spectra
object is determined
(taking into account all the MS levels),
(2) the Spectra
object is filtered according to the MS level and
subsequently ordered according to the retention time,
(3) the MS events are split into four (approximately) equal parts,
(4) the relative retention time is calculated (using the retention time
duration from (1) and taking into account the minimum retention time),
(5) the relative retention time values associated to the MS event parts
are returned.
- rtOverMsQuarters, MS2 quarter RT fraction (MS:4000056),
"The interval used for acquisition of the first, second, third, and fourth
quarter of all MS2 events divided by retention time duration." [PSI:MS],
msLevel = 2L
;
The metric informs about the dynamic range of the acquisition along the
chromatographic separation. For MS2 scans, the comparability of the values
depends on the acquisition mode and settings to select ions for fragmentation.
The metric is calculated as follows:
(1) the retention time duration of the whole Spectra
object is determined
(taking into account all the MS levels),
(2) the Spectra
object is filtered according to the MS level and
subsequently ordered according to the retention time,
(3) the MS events are split into four (approximately) equal parts,
(4) the relative retention time is calculated (using the retention time
duration from (1) and taking into account the minimum retention time),
(5) the relative retention time values associated to the MS event parts
are returned.
- ticQuartileToQuartileLogRatio, MS1 TIC-change quartile ratios
(MS:4000057), ""The log ratios of successive TIC-change quartiles. The TIC
changes are the list of MS1 total ion current (TIC) value changes from
one to the next scan, produced when each MS1 TIC is subtracted from the
preceding MS1 TIC. The metric's value triplet represents the log ratio of the
TIC-change Q2 to Q1, Q3 to Q2, TIC-change-max to Q3" [PSI:MS],
mode = "TIC_change"
, relativeTo = "previous"
, msLevel = 1L
;
The metric informs about the dynamic range of the acquisition along the
chromatographic separation.This metric evaluates the stability (similarity)
of MS1 TIC values from scan to scan along the LC run. High log ratios
representing very large intensity differences between pairs of scans might
be due to electrospray instability or presence of a chemical contaminant.
The metric is calculated as follows:
(1) the TIC (ionCount
) of the Spectra
object is calculated
per scan event (with spectra ordered by retention time),
(2) the differences between TIC values are calculated between subsequent
scan events,
(3) the ratios between the 25\%, 50\%, 75\%, and 100\% quantile to the
25\% quantile of the values of (2) are calculated,
(4) the log
values of the ratios are returned.
- ticQuartileToQuartileLogRatio, MS1 TIC quartile ratios (MS:4000058),
"The log ratios of successive TIC quartiles. The metric's value triplet
represents the log ratios of TIC-Q2 to TIC-Q1, TIC-Q3 to TIC-Q2,
TIC-max to TIC-Q3." [PSI:MS], mode = "TIC"
,
relativeTo = "previous"
, msLevel = 1L
;
The metric informs about the dynamic range of the acquisition along the
chromatographic separation. The ratios provide information on the distribution
of the TIC values for one LC-MS run. Within an experiment, with the same
LC setup, values should be comparable between samples.
The metric is calculated as follows:
(1) the TIC (ionCount
) of the Spectra
object is calculated
per scan event (with spectra ordered by retention time),
(2) the TIC values between subsequent scan events are taken
as they are,
(3) the ratios between the 25\%, 50\%, 75\%, and 100\% quantile to the
25\% quantile of the values of (2) are calculated.
(4) The log
values of the ratios are returned.
- numberSpectra, number of MS1 spectra MS:4000059),
"The number of MS1 events in the run." [PSI:MS], msLevel = 1L
;
An unusual low number may indicate incomplete sampling/scan rate of the MS
instrument, low sample volume and/or failed injection of a sample.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the number of the spectra are obtained (length
of Spectra
) and
returned.
- numberSpectra, number of MS2 spectra (MS:4000060),
"The number of MS2 events in the run." [PSI:MS], msLevel = 2L
;
An unusual low number may indicate incomplete sampling/scan rate of the MS
instrument, low sample volume and/or failed injection of a sample.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the number of the spectra are obtained (length
of Spectra
) and
returned.
- mzAcquisitionRange, m/z acquisition range (MS:4000069),
"Upper and lower limit of m/z precursor values at which MSn spectra are
recorded." [PSI:MS];
The metric informs about the dynamic range of the acquisition. Based on the
used MS instrument configuration, the values should be similar. Variations
between measurements may arise when employing acquisition in DDA mode.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the m/z values of the peaks within the Spectra
object are obtained,
(3) the minimum and maximum m/z values are obtained and returned.
- rtAcquisitionRange, retention time acquisition range (MS:4000070),
"Upper and lower limit of retention time at which spectra are recorded."
[PSI:MS];
An unusual low range may indicate incomplete sampling and/or a premature or
failed LC run.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the retention time values of the features within the Spectra
object are obtained,
(3) the minimum and maximum retention time values are obtained and
returned.
- msSignal10xChange, MS1 signal jump (10x) count (MS:4000097),
"The number of times where MS1 TIC increased more than 10-fold between adjacent
MS1 scans. An unusual high count of signal jumps or falls can indicate
ESI stability issues." [PSI:MS], change = "jump"
, msLevel = 1L
;
An unusual high count of signal jumps or falls may indicate ESI stability
issues.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensity of the precursor ions within the Spectra
object are
obtained,
(3) the intensity values of the features are obtained via the ion count,
(4) the signal jumps/declines of the intensity values with the two
subsequent intensity values is calculated,
(5) the signal jumps by a factor of ten or more are counted and returned.
- msSignal10xChange, MS1 signal fall (10x) count (MS:4000098),
"The number of times where MS1 TIC decreased more than 10-fold between adjacent
MS1 scans. An unusual high count of signal jumps or falls can indicate
ESI stability issues." [PSI:MS], change = "fall"
, msLevel = 1L
;
An unusual high count of signal jumps or falls may indicate ESI stability
issues.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensity of the precursor ions within the Spectra
object are
obtained,
(3) the intensity values of the features are obtained via the ion count,
(4) the signal jumps/declines of the intensity values with the two
subsequent intensity values is calculated,
(5) the signal declines by a factor of ten or more are counted and returned.
- numberEmptyScans, number of empty MS1 scans (MS:4000099),
"Number of MS1 scans where the scans' peaks intensity sums to 0 (i.e. no
peaks or only 0-intensity peaks)." [PSI:MS], msLevel = 1L
;
An unusual high number may indicate incomplete sampling/scan rate of the MS
instrument, low sample volume and/or failed injection of a sample.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensities per entry are obtained,
(3) the number of intensity entries that are NULL
, NA
, or
that have a sum of 0
are obtained and returned.
- numberEmptyScans, number of empty MS2 scans (MS:4000100),
"Number of MS2 scans where the scans' peaks intensity sums to 0 (i.e. no
peaks or only 0-intensity peaks)." [PSI:MS], msLevel = 2L
;
An unusual high number may indicate incomplete sampling/scan rate of the MS
instrument, low sample volume and/or failed injection of a sample.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensities per entry are obtained,
(3) the number of intensity entries that are NULL
, NA
, or
that have a sum of 0
are obtained and returned.
- numberEmptyScans, number of empty MS3 scans (MS:4000101),
"Number of MS3 scans where the scans' peaks intensity sums to 0 (i.e. no
peaks or only 0-intensity peaks)." [PSI:MS], msLevel = 3L
;
An unusual high number may indicate incomplete sampling/scan rate of the MS
instrument, low sample volume and/or failed injection of a sample.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensities per entry are obtained,
(3) the number of intensity entries that are NULL
, NA
, or
that have a sum of 0
are obtained and returned.
- precursorIntensityQuartiles,
MS2 precursor intensity distribution Q1, Q2, Q3 (MS:4000116),
"From the distribution of MS2 precursor intensities, the quartiles
Q1, Q2, Q3." [PSI:MS], identificationLevel = "all"
;
The intensity distribution of the precursors informs about the dynamic range
of the acquisition.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensity of the precursor ions within the Spectra
object are obtained,
(3) the 25\%, 50\%, and 75\% quantile of the precursor intensity values are
obtained (NA
values are removed) and returned.
- precursorIntensityMean, MS2 precursor intensity distribution mean
(MS:4000117), "From the distribution of MS2 precursor intensities, the mean."
[PSI:MS], identificationLevel = "all"
;
The intensity distribution of the precursors informs about the dynamic range
of the acquisition.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensity of the precursor ions within the Spectra
object are
obtained,
(3) the mean of the precursor intensity values is obtained
(NA
values are removed) and returned.
- precursorIntensitySd, MS2 precursor intensity distribution sigma
(MS:4000118), "From the distribution of MS2 precursor intensities, the sigma
value." [PSI:MS], identificationLevel = "all"
;
The intensity distribution of the precursors informs about the dynamic range of
the acquisition.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensity of the precursor ions within the Spectra
object are
obtained,
(3) the standard deviation of precursor intensity values is obtained
(NA
values are removed) and returned.
- medianPrecursorMz,
MS2 precursor median m/z of identified quantification data points
(MS:4000152),
"Median m/z value for MS2 precursors of all quantification data points after
user-defined acceptance criteria are applied. These data points may be for
example XIC profiles, isotopic pattern areas, or reporter ions
(see MS:1001805). The used type should be noted in the metadata or
analysis methods section of the recording file for the respective run. In
case of multiple acceptance criteria (FDR) available in proteomics, PSM-level
FDR should be used for better comparability." [PSI:MS],
identificationLevel = "identified"
, msLevel = 1L
;
The m/z distribution informs about the dynamic range of the acquisition.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the precursor m/z values are obtained,
(3) the median value is returned (NA
s are removed).
- rtIqr,
interquartile RT period for identified quantification data points
(MS:4000153), "The interquartile retention time period, in seconds, for all
quantification data points after user-defined acceptance criteria are applied
over the complete run. These data points may be for example XIC profiles,
isotopic pattern areas, or reporter ions (see MS:1001805). The used type
should be noted in the metadata or analysis methods section of the recording
file for the respective run. In case of multiple acceptance criteria (FDR)
available in proteomics, PSM-level FDR should be used for better
comparability."
[PSI:MS], identificationLevel = "identified"
;
Longer duration may indicate a better chromatographic separation of compounds
which depends, however, also on the sampling/scan rate of the MS instrument.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the retention time values are obtained,
(3) the interquartile range is obtained from the values and returned
(NA
values are removed).
- rtIqrRate,
rate of the interquartile RT period for identified quantification data points
(MS:4000154), "The rate of identified quantification data points for the
interquartile retention time period, in identified quantification data points
per second. These data points may be for example XIC profiles, isotopic
pattern areas, or reporter ions (see MS:1001805). The used type should
be noted in the metadata or analysis methods section of the recording
file for the respective run. In case of multiple acceptance criteria (FDR)
available in proteomics, PSM-level FDR should be used for better
comparability." [PSI:MS],
identificationLevel = "identified"
;
Higher rates may indicate a more efficient sampling and identification.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the retention time values are obtained,
(3) the 25\% and 75\% quantiles are obtained from the retention time values
(NA
values are removed),
(4) the number of eluted features between this 25\% and 75\% quantile is
calculated,
(5) the number of features is divided by the interquartile range of the
retention time and returned.
- areaUnderTic, area under TIC (MS:4000155),
"The area under the total ion chromatogram." [PSI:MS];
The metric informs about the dynamic range of the acquisition. Differences
between samples of an experiment may indicate differences in the dynamic
range and/or in the sample content.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the sum of the ion counts are obtained and returned.
- areaUnderTicRtQuantiles, area under TIC RT quantiles (MS:4000156),
"The area under the total ion chromatogram of the retention time quantiles.
Number of quantiles are given by the n-tuple." [PSI:MS];
The metric informs about the dynamic range of the acquisition. Differences
between samples of an experiment may indicate differences in the dynamic
range and/or in the sample content.
The metric informs about the dynamic range of the acquisition along the
chromatographic separation. Differences between samples of an experiment may
indicate differences in chromatographic performance, differences in the
dynamic range and/or in the sample content.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the Spectra
object is ordered according to the retention time,
(3) the 0\%, 25\%, 50\%, 75\%, and 100\% quantiles of the retention time
values are obtained,
(4) the ion count of the intervals between the 0\%/25\%, 25\%/50\%,
50\%/75\%, and 75\%/100\% are obtained, \
(5) the ion counts of the intervals are summed (TIC) and the values returned.
- extentIdentifiedPrecursorIntensity,
extent of identified MS2 precursor intensity (MS:4000157),
"Ratio of 95th over 5th percentile of MS2 precursor intensity for all
quantification data points after user-defined acceptance criteria are
applied. The used type of identification should be noted in the metadata or
analysis methods section of the recording file for the respective run.
In case of multiple acceptance criteria (FDR) available in proteomics,
PSM-level FDR should be used for better comparability." [PSI:MS],
identificationLevel = "identified"
;
The metric informs about the dynamic range of the acquisition.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensities of the precursor ions are obtained,
(3) the 5\% and 95\% quantile of these intensities are obtained
(1NA1 values are removed),
(4) the ratio between the 95\% and the 5\% intensity quantile is calculated
and returned.
- medianTicRtIqr, median of TIC values in the RT range in which the middle
half of quantification data points are identified (MS:4000158),
"Median of TIC values in the RT range in which half of quantification data
points are identified (RT values of Q1 to Q3 of identifications). These
data points may be for example XIC profiles, isotopic pattern areas, or
reporter ions (see MS:1001805). The used type should be noted in the metadata
or analysis methods section of the recording file for the respective run.
In case of multiple acceptance criteria (FDR) available in proteomics,
PSM-level FDR should be used for better comparability." [PSI:MS],
identificationLevel = "identified"
;
The metric informs about the dynamic range of the acquisition along the
chromatographic separation.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the Spectra
object is ordered according to the retention time,
(3) the features between the 1st and 3rd quartile are obtained
(half of the features that are present in the Spectra
object),
(4) the ion count of the features within the 1st and 3rd quartile is
obtained,
(5) the median value of the ion count is calculated (NA
values are
removed) and the median value is returned.
- medianTicOfRtRange, median of TIC values in the shortest RT range in
which half of the quantification data points are identified (MS:4000159),
"Median of TIC values in the shortest RT range in which half of the
quantification data points are identified. These data points may be for
example XIC profiles, isotopic pattern areas, or reporter ions
(see MS:1001805). The used type should be noted in the metadata or analysis
methods section of the recording file for the respective run. In case of
multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR
should be used for better comparability." [PSI:MS],
identificationLevel = "identified"
;
The metric informs about the dynamic range of the acquisition along the
chromatographic separation.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the Spectra
object is ordered according to the retention time,
(3) the number of features in the Spectra
object is obtained and
the number for half of the features is calculated,
(4) iterate through the features (always by taking the neighbouring
half of features) and calculate the retention time range of the
set of features,
(5) retrieve the set of features with the minimum retention time
range,
(6) calculate from the set of (5) the median TIC (NA
values are removed)
and return it.
- precursorIntensityRange, MS2 precursor intensity range (MS:4000160),
"Minimum and maximum MS2 precursor intensity recorded. " [PSI:MS];
The metric informs about the dynamic range of the acquisition.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensity of the precursor ions within the Spectra
object are
obtained,
(3) the minimum and maximum precursor intensity values are obtained and
returned.
- precursorIntensityQuartiles,
identified MS2 precursor intensity distribution Q1, Q2, Q3 (MS:4000161),
"From the distribution of identified MS2 precursor intensities, the quartiles
Q1, Q2, Q3. The used type of identification should be noted in the metadata
or analysis methods section of the recording file for the respective run.
In case of multiple acceptance criteria (FDR) available in proteomics,
PSM-level FDR should be used for better comparability." [PSI:MS],
identificationLevel = "identified"
;
The metric informs about the dynamic range of the acquisition in relation to
identifiability.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensity of the precursor ions within the Spectra
object are obtained,
(3) the 25\%, 50\%, and 75\% quantile of the precursor intensity values are
obtained (NA
values are removed) and returned.
- precursorIntensityQuartiles,
unidentified MS2 precursor intensity distribution Q1, Q2, Q3 (MS:4000162),
"From the distribution of unidentified MS2 precursor intensities, the
quartiles Q1, Q2, Q3. The used type of identification should be noted in the
metadata or analysis methods section of the recording file for the respective
run. In case of multiple acceptance criteria (FDR) available in proteomics,
PSM-level FDR should be used for better comparability."
[PSI:MS], identificationLevel = "unidentified"
;
The metric informs about the dynamic range of the acquisition in relation to
identifiability.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensity of the precursor ions within the Spectra
object are
obtained,
(3) the 25\%, 50\%, and 75\% quantile of the precursor intensity values are
obtained (NA
values are removed) and returned.
- precursorIntensityMean,
identified MS2 precursor intensity distribution mean (MS:4000163),
"From the distribution of identified MS2 precursor intensities, the mean.
The intensity distribution of the identified precursors informs about the
dynamic range of the acquisition in relation to identifiability. The used
type of identification should be noted in the metadata or analysis methods
section of the recording file for the respective run. In case of multiple
acceptance criteria (FDR) available in proteomics, PSM-level FDR should be
used for better comparability."
[PSI:MS], identificationLevel = "identified"
;
The metric informs about the dynamic range of the acquisition in relation to
identifiability.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensity of the precursor ions within the Spectra
object are
obtained,
(3) the mean of the precursor intensity values is obtained
(NA
values are removed) and returned.
- precursorIntensityMean,
unidentified MS2 precursor intensity distribution mean (MS:4000164),
"From the distribution of unidentified MS2 precursor intensities, the mean.
The used type of identification should be noted in the metadata or analysis
methods section of the recording file for the respective run. In case of
multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR
should be used for better comparability." [PSI:MS],
identificationLevel = "unidentified"
;
The metric informs about the dynamic range of the acquisition in relation
to identifiability.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensity of the precursor ions within the Spectra
object are
obtained,
(3) the mean of the precursor intensity values is obtained
(NA
values are removed) and returned.
- precursorIntensitySd,
identified MS2 precursor intensity distribution sigma (MS:4000165),
"From the distribution of identified MS2 precursor intensities, the sigma
value. The used type of identification should be noted in the metadata or
analysis methods section of the recording file for the respective run. In
case of multiple acceptance criteria (FDR) available in proteomics, PSM-level
FDR should be used for better comparability." [PSI:MS],
identificationLevel = "identified"
;
The metric informs about the dynamic range of the acquisition in relation to
identifiability.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensity of the precursor ions within the Spectra
object are
obtained,
(3) the standard deviation of precursor intensity values is obtained
(NA
values are removed) and returned.
- precursorIntensitySD,
unidentified MS2 precursor intensity distribution sigma (MS:4000166),
"From the distribution of unidentified MS2 precursor intensities, the sigma
value. The used type of identification should be noted in the metadata or
analysis methods section of the recording file for the respective run. In
case of multiple acceptance criteria (FDR) available in proteomics, PSM-level
FDR should be used for better comparability." [PSI:MS],
identificationLevel = "unidentified"
;
The metric informs about the dynamic range of the acquisition in relation to
identifiability.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the intensity of the precursor ions within the Spectra
object are
obtained,
(3) the standard deviation of precursor intensity values is obtained
(NA
values are removed) and returned.
- ratioCharge1over2,
ratio of 1+ over 2+ of all MS2 known precursor charges (MS:4000167),
"The ratio of 1+ over 2+ MS2 precursor charge count of all spectra." [PSI:MS],
identificationLevel = "all"
;
High ratios of 1+/2+ MS2 precursor charge count may indicate inefficient
ionization.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the precursor charge is obtained,
(3) the number of precursors with charge 1+ is divided by the number of
precursors with charge 2+ and the ratio is returned.
- ratioCharge1over2,
ratio of 1+ over 2+ of identified MS2 known precursor charges (MS:4000168),
""The ratio of 1+ over 2+ MS2 precursor charge count of identified spectra.
The used type of identification should be noted in the metadata or analysis
methods section of the recording file for the respective run. In case of
multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR
should be used for better comparability." [PSI:MS],
identificationLevel = "identified"
;
High ratios of 1+/2+ MS2 precursor charge count may indicate inefficient
ionization in relation to identifiability.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the precursor charge is obtained,
(3) the number of precursors with charge 1+ is divided by the number of
precursors with charge 2+ and the ratio is returned.
- ratioCharge3over2,
ratio of 3+ over 2+ of all MS2 known precursor charges (MS:4000169),
"The ratio of 3+ over 2+ MS2 precursor charge count of all spectra." [PSI:MS],
identificationLevel = "all"
;
Higher ratios of 3+/2+ MS2 precursor charge count may indicate e.g.
preference for longer peptides.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the precursor charge is obtained,
(3) the number of precursors with charge 3+ is divided by the number of
precursors with charge 2+ and the ratio is returned.
- ratioCharge3over2,
ratio of 3+ over 2+ of identified MS2 known precursor charges (MS:4000170),
"The ratio of 3+ over 2+ MS2 precursor charge count of identified spectra.
The used type of identification should be noted in the metadata or analysis
methods section of the recording file for the respective run. In case of
multiple acceptance criteria (FDR) available in proteomics, PSM-level
FDR should be used for better comparability." [PSI:MS],
identificationLevel = "identified"
;
Higher ratios of 3+/2+ MS2 precursor charge count may indicate e.g.
preference for longer peptides in relation to identifiability.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the precursor charge is obtained,
(3) the number of precursors with charge 3+ is divided by the number of
precursors with charge 2+ and the ratio is returned.
- ratioCharge4over2,
ratio of 4+ over 2+ of all MS2 known precursor charges (MS:4000171),
"The ratio of 4+ over 2+ MS2 precursor charge count of all spectra."
[PSI:MS], identificationLevel = "all"
;
Higher ratios of 3+/2+ MS2 precursor charge count may indicate e.g.
preference for longer peptides.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the precursor charge is obtained,
(3) the number of precursors with charge 4+ is divided by the number of
precursors with charge 2+ and the ratio is returned.
- ratioCharge4over2,
ratio of 4+ over 2+ of identified MS2 known precursor charges (MS:4000172),
"The ratio of 4+ over 2+ MS2 precursor charge count of identified spectra.
The used type of identification should be noted in the metadata or analysis
methods section of the recording file for the respective run. In case of
multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR
should be used for better comparability." [PSI:MS],
identificationLevel = "identified"
;
Higher ratios of 3+/2+ MS2 precursor charge count may indicate e.g.
preference for longer peptides in relation to identifiability.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the precursor charge is obtained,
(3) the number of precursors with charge 4+ is divided by the number of
precursors with charge 2+ and the ratio is returned.
- meanCharge, mean MS2 precursor charge in all spectra (MS:4000173),
"Mean MS2 precursor charge in all spectra" [PSI:MS],
identificationLevel = "all"
;
Higher charges may indicate inefficient ionization or e.g. preference for
longer peptides.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the precursor charge is obtained,
(3) the mean of the precursor charge values is calculated and returned.
- meanCharge, mean MS2 precursor charge in identified spectra (MS:4000174),
"Mean MS2 precursor charge in identified spectra. The used type of
identification should be noted in the metadata or analysis methods section
of the recording file for the respective run. In case of multiple acceptance
criteria (FDR) available in proteomics, PSM-level FDR should be used for
better comparability." [PSI:MS], identificationLevel = "identified"
;
Higher charges may indicate inefficient ionization or e.g. preference for
longer peptides in relation to identifiability.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the precursor charge is obtained,
(3) the mean of the precursor charge values is calculated and returned.
- medianCharge, median MS2 precursor charge in all spectra (MS:4000175),
"Median MS2 precursor charge in all spectra" [PSI:MS],
identificationLevel = "all"
;
Higher charges may indicate inefficient ionization and/or e.g. preference
for longer peptides.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the precursor charge is obtained,
(3) the median of the precursor charge values is calculated and returned.
- medianCharge, median MS2 precursor charge in identified spectra
(MS:4000176), "Median MS2 precursor charge in identified spectra. The used
type of identification should be noted in the metadata or analysis methods
section of the recording file for the respective run. In case of multiple
acceptance criteria (FDR) available in proteomics, PSM-level FDR should be
used for better comparability." [PSI:MS],
identificationLevel = "identified"
;
Higher charges may indicate inefficient ionization and/or e.g. preference for
longer peptides in relation to identifiability.
The metric is calculated as follows:
(1) the Spectra
object is filtered according to the MS level,
(2) the precursor charge is obtained,
(3) the median of the precursor charge values is calculated and returned.
- ticQuantileRtFraction, TIC quantile RT fraction (MS:4000183),
"The interval when the respective quantile of the TIC accumulates divided by
retention time duration. The number of values in the tuple implies the
quantile mode." [PSI:MS];
The metric informs about the dynamic range of the acquisition along the
chromatographic separation. The metric provides information on the sample
(compound) flow along the chromatographic run, potentially revealing poor
chromatographic performance, such as the absence of a signal for a
significant portion of the run.
The metric is calculated as follows:
(1) the Spectra
object is ordered according to the retention time,
(2) the cumulative sum of the ion count is calculated (TIC),
(3) the quantiles are calculated according to the probs
argument,
e.g. when probs
is set to c(0, 0.25, 0.5, 0.75, 1)
the
0\%, 25\%, 50\%, 75\%, and 100\% quantile is calculated,
(4) the retention time/relative retention time (retention time divided by
the total run time taking into account the minimum retention time) is
calculated,
(5) the (relative) duration of the LC run after which the cumulative
TIC exceeds (for the first time) the respective quantile of the
cumulative TIC is calculated and returned.
data.frame
-formatThe most important function to assess the data quality and to calculate the
metrics is the calculateMetrics
function. The function takes
a Spectra
or MsExperiment
object as input, a character vector of metrics
to be calculated, a Boolean value to the filterEmptySpectra
argument,
and, optionally a list of parameters passed to the quality metrics functions.
The filterEmptySpectra
argument specifies if zero-intensity, Inf
-intensity
or zero-length entries should be removed (filterEmptySpectra = TRUE
).
By default, the entries are taken as they are (filterEmptySpectra = FALSE
).
The argument can be set to TRUE
to compute metrics that are close to the
implementation of the QuaMeter
software. Prior to calculating the metrics,
the implementation of QuaMeter
skips all spectra with defaultArrayLength=0
(in .mzML files) at any MS level.
When passing a Spectra
/MsExperiment
object to the function, a data.frame
returned by calculateMetrics
with the metrics specified by the argument
metrics
. By default, qualityMetrics(object)
is taken to specify
the calculation of quality metrics. calculateMetrics
also accepts a list
of parameters passed to the individual quality metrics functions. For
each quality metrics functions, the relevant parameters are selected based on
the accepted arguments.
Additional arguments can be given to the quality metrics functions.
For example, the function ticQuartileToQuartileLogRatio
function has the
arguments relativeTo
, mode
, and msLevel
. relativeTo
specifies to which
quantile the log TIC quantile is relatively related to (either to the 1st
quantile or the respective previous one). mode
(either "TIC_change"
or
"TIC"
) specifies if the quantiles are taken from the changes between TICs
of scan events or the TICs directly. One Spectra
/MsExperiment
object may
also contain more than one msLevel
, e.g. if it also contains information on
MS$^2$ or MS$^3$ features. If the user adds the arguments
relativeTo = "Q1", mode = "TIC", msLevel = c(1L, 2L))
,
ticQuartileToQuartileLogRatio
is run with the parameter combinations
relativeTo = "Q1", mode = "TIC", msLevel = c(1L, 2L)
.
The results based on these parameter combinations are returned and the used parameters are returned as attributes to the returned vector.
Here, we would like to calculate the metrics of all included quality
metrics functions (qualityMetrics(object)
) and additionally pass the
parameter relativeTo = "Q1"
and relativeTo = "previous"
. For computational
reasons, we will restrict the calculation of the metrics to the first
sample and to RPLC samples.
## subset the Spectra objects sps_comb_subset <- sps_comb[grep("Sample.1_", sps_comb$dataOrigin), ] ## for RPLC and HILIC metrics_sps_Q1 <- calculateMetrics(object = sps_comb_subset, metrics = qualityMetrics(sps_comb_subset), filterEmptySpectra = FALSE, relativeTo = "Q1", msLevel = 1L) metrics_sps_Q1 metrics_sps_previous <- calculateMetrics(object = sps_comb_subset, metrics = qualityMetrics(sps_comb_subset), filterEmptySpectra = FALSE, relativeTo = "previous", msLevel = 1L) metrics_sps_previous
Alternatively, an MsExperiment
object might be passed to
calculateMetrics
. The function will iterate over the samples (referring
to rows in sampleData(msexp))
) and calculate the quality metrics on the
corresponding Spectra
s.
mzQC
-formatBy default, a data.frame
object containing the metric values as entries
are returned by the the function calculateMetrics
. Alternatively, the
function also allows the user to export the metrics in a format defined by
the rmzqc
package by setting the argument format
to "mzQC"
(default:
format = "data.frame"
). In that case, only the metrics that comply to the
mzQC
specification will be written to the returned object.
The object can be exported and validated using the functionality of the rmzqc
package (see the documentation of rmzqc
for further information).
There are in total r length(msexp_hilic)
samples
respectively in the objects msexp_rplc
and msexp_hilic
. To improve
the visualization and interpretability, we will only calculate the metrics
from the first 20 of these samples.
In this example here, we will remove zero-length and zero-intensity entries
prior to calculating the metrics. To do this, we set the filterEmptySpectra
argument to TRUE
within the calculateMetrics
function.
## subset the MsExperiment objects msexp_rplc_subset <- msexp_rplc[1:20] msexp_hilic_subset <- msexp_hilic[1:20] ## define metrics metrics_sps <- c("chromatographyDuration", "ticQuantileRtFraction", "rtOverMsQuarters", "ticQuartileToQuartileLogRatio", "numberSpectra", "medianPrecursorMz", "rtIqr", "rtIqrRate", "areaUnderTic") ## for RPLC-derived MsExperiment metrics_rplc_msexp <- calculateMetrics(object = msexp_rplc_subset, metrics = qualityMetrics(msexp_rplc_subset), filterEmptySpectra = TRUE, relativeTo = "Q1", msLevel = 1L) ## for HILIC-derived MsExperiment metrics_hilic_msexp <- calculateMetrics(object = msexp_hilic_subset, metrics = qualityMetrics(msexp_hilic_subset), filterEmptySpectra = TRUE, relativeTo = "Q1", msLevel = 1L)
When passing an MsExperiment
object to calculateMetrics
a data.frame
object is returned with the samples (derived from the rownames of
sampleData(msexp)
) in the rows and the metrics in columns.
We will show here the objects metrics_rplc_msexp
and metrics_hilic_msexp
print("metrics_rplc_msexp") rmarkdown::paged_table(as.data.frame(metrics_rplc_msexp)) print("metrics_hilic_msexp") rmarkdown::paged_table(as.data.frame(metrics_hilic_msexp))
The quality metrics can be most easily compared when graphically visualized.
The MsQuality
package offers the possibility to graphically display the
metrics using the plotMetric
and shinyMsQuality
functions. The
plotMetric
function will create one plot based on a single metric.
shinyMsQuality
, on the other hand, opens a shiny application that allows
to browse through all the metrics stored in the object.
As a way of example, we will plot here the number of features. A high number of missing features might indicate low data quality, however, also different sample types might exhibit contrasting number of detected features. As a general rule, only samples of the same type should be compared to adjust for sample type-specific effects.
metrics_msexp <- rbind(metrics_rplc_msexp, metrics_hilic_msexp) plotMetric(qc = metrics_msexp, metric = "numberSpectra")
Similarly, we are able to display the area under the TIC for the retention time quantiles. This plot gives information on the perceived signal (TIC) for the differnt retention time quantiles and could indicate drifts or interruptions of sensitivity during the run.
plotMetric(qc = metrics_msexp, metric = "ticQuartileToQuartileLogRatio")
Alternatively, to browse through all metrics that were calculated in an
interactive way, we can use the shinyMsQuality
function.
shinyMsQuality(qc = metrics_msexp)
All software and respective versions to build this vignette are listed here:
sessionInfo()