# @RdocDocumentation "1. Calibration and Normalization"
#
# \encoding{latin1}
#
# \description{
# In this section we give \emph{our} recommendation on how spotted
# two-color (or multi-color) microarray data is best calibrated and
# normalized.
# }
#
# \section{Classical background subtraction}{
# We do \emph{not} recommend background subtraction in classical
# means where background is estimated by various image analysis
# methods. This means that we will only consider foreground signals
# in the analysis.
#
# We estimate "background" by other means. In what is explain below,
# only a global background, that is, a global bias, is estimated
# and removed.
# }
#
# \section{Multiscan calibration}{
# In Bengtsson et al (2004) we give evidence that microarray scanners
# can introduce a significant bias in data. This bias, which is
# about 15-25 out of 65535, \emph{will} introduce intensity dependency
# in the log-ratios, as explained in Bengtsson &
# \enc{Hössjer}{Hossjer} (2006).
#
# In Bengtsson et al (2004) we find that this bias is stable across
# arrays (and a couple of months), but further research is needed
# in order to tell if this is true over a longer time period.
#
# To calibrate signals for scanner biases, scan the same array at
# multiple PMT-settings at three or more (K >= 3) different
# PMT settings (preferably in decreasing order).
# While doing this, \emph{do not adjust the laser power settings}.
# Also, do the multiscan \emph{without} washing, cleaning or by other
# means changing the array between subsequent scans.
# Although not necessary, it is preferred that the array
# remains in the scanner between subsequent scans. This will simplify
# the image analysis since spot identification can be made once
# if images aligns perfectly.
#
# After image analysis, read all K scans for the same array into the
# two matrices, one for the red and one for the green channel, where
# the K columns corresponds to scans and the N rows to the spots.
# It is enough to use foreground signals.
#
# In order to multiscan calibrate the data, for each channel
# separately call \code{Xc <- calibrateMultiscan(X)} where \code{X}
# is the NxK matrix of signals for one channel across all scans. The
# calibrated signals are returned in the Nx1 matrix \code{Xc}.
#
# Multiscan calibration may sometimes be skipped, especially if affine
# normalization is applied immediately after, but we do recommend that
# every lab check at least once if their scanner introduce bias.
# If the offsets in a scanner is already estimated from earlier
# multiscan analyses, or known by other means, they can readily be
# subtracted from the signals of each channel. If arrays are still
# multiscanned, it is possible to force the calibration method to
# fit the model with zero intercept (assuming the scanner offsets
# have been subtracted) by adding argument \code{center=FALSE}.
# }
#
# \section{Affine normalization}{
# In Bengtsson & \enc{Hössjer}{Hossjer} (2006), we carry out a detailed
# study on how biases in each channel introduce so called
# intensity-dependent log-ratios among other systematic artifacts.
# Data with (additive) bias in each channel is said to be \emph{affinely}
# transformed. Data without such bias, is said to be \emph{linearly}
# (proportionally) transform. Ideally, observed signals (data) is a
# linear (proportional) function of true gene expression levels.
#
# We do \emph{not} assume proportional observations. The scanner bias
# is real evidence that assuming linearity is not correct.
# Affine normalization corrects for affine transformation in data.
# Without control spots it is not possible to estimate the bias in each
# of the channels but only the relative bias such that after
# normalization the effective bias are the same in all channels.
# This is why we call it normalization and not calibration.
#
# In its simplest form, affine normalization is done by
# \code{Xn <- normalizeAffine(X)} where \code{X} is a Nx2 matrix with
# the first column holds the foreground signals from the red channel and
# the second holds the signals from the green channel. If three- or
# four-channel data is used these are added the same way. The normalized
# data is returned as a Nx2 matrix \code{Xn}.
#
# To normalize all arrays and all channels at once, one may put all
# data into one big NxK matrix where the K columns hold the all channels
# from the first array, then all channels from the second array and so
# on. Then \code{Xn <- normalizeAffine(X)} will return the across-array
# and across-channel normalized data in the NxK matrix \code{Xn} where
# the colunms are stored in the same order as in matrix \code{X}.
#
# Equal effective bias in all channels is much better. First of all,
# any intensity-dependent bias in the log-ratios is removed \emph{for
# all non-differentially expressed genes}. There is still an
# intensity-dependent bias in the log-ratios for differentially expressed
# genes, but this is now symmetric around log-ratio zero.
#
# Affine normalization will (by default and recommended) normalize
# \emph{all} arrays together and at once. This will guarantee that
# all arrays are "on the same scale". Thus, it \emph{not} recommended
# to apply a classical between-array scale normalization afterward.
# Moreover, the average log-ratio will be zero after an affine
# normalization.
#
# Note that an affine normalization will only remove curvature in the
# log-ratios at lower intensities.
# If a strong intensity-dependent bias at high intensities remains,
# this is most likely due to saturation effects, such as too high PMT
# settings or quenching.
#
# Note that for a perfect affine normalization you \emph{should}
# expect much higher noise levels in the \emph{log-ratios} at lower
# intensities than at higher. It should also be approximately
# symmetric around zero log-ratio.
# In other words, \emph{a strong fanning effect is a good sign}.
#
# Due to different noise levels in red and green channels, different
# PMT settings in different channels, plus the fact that the
# minimum signal is zero, "odd shapes" may be seen in the log-ratio
# vs log-intensity graphs at lower intensities. Typically, these
# show themselves as non-symmetric in positive and negative log-ratios.
# Note that you should not see this at higher intensities.
#
# If there is a strong intensity-dependent effect left after the
# affine normalization, we recommend, for now, that a subsequent
# curve-fit or quantile normalization is done.
# Which one, we do not know.
#
# Why negative signals?
# By default, 5\% of the normalized signals will have a non-positive
# signal in one or both channels. \emph{This is on purpose}, although
# the exact number 5\% is chosen by experience. The reason for
# introducing negative signals is that they are indeed expected.
# For instance, when measure a zero gene expression level, there is
# a chance that the observed value is (should be) negative due to
# measurement noise. (For this reason it is possible that the scanner
# manufacturers have introduced scanner bias on purpose to avoid
# negative signals, which then all would be truncated to zero.)
# To adjust the ratio (or number) of negative signals allowed, use
# for example \code{normalizeAffine(X, constraint=0.01)} for 1\%
# negative signals. If set to zero (or \code{"max"}) only as much
# bias is removed such that no negative signals exist afterward.
# Note that this is also true if there were negative signals on
# beforehand.
#
# Why not lowess normalization?
# Curve-fit normalization methods such as lowess normalization are
# basically designed based on linearity assumptions and will for this
# reason not correct for channel biases. Curve-fit normalization
# methods can by definition only be applied to one pair of channels
# at the time and do therefore require a subsequent between-array
# scale normalization, which is by the way very ad hoc.
#
# Why not quantile normalization?
# Affine normalization can be though of a special case of quantile
# normalization that is more robust than the latter.
# See Bengtsson & \enc{Hössjer}{Hossjer} (2006) for details.
# Quantile normalization is probably better to apply than curve-fit
# normalization methods, but less robust than affine normalization,
# especially at extreme (low and high) intensities.
# For this reason, we do recommend to use affine normalization first,
# and if this is not satisfactory, quantile normalization may be applied.
# }
#
# \section{Linear (proportional) normalization}{
# If the channel offsets are zero, already corrected for, or estimated
# by other means, it is possible to normalize the data robustly by
# fitting the above affine model without intercept, that is, fitting
# a truly linear model. This is done adding argument \code{center=FALSE}
# when calling \code{normalizeAffine()}.
# }
#
# @author "HB"
#*/#########################################################################
############################################################################
# HISTORY:
# 2011-02-05
# o DOCUMENTATION: Added paragraphs on how to do affine normalization
# when channel offsets are known/zero. Same for multiscan calibration
# when scanner offsets are known/zero.
# 2006-06-29
# o Added to aroma.light instead.
# 2005-02-02
# o Created.
############################################################################